spegg © VIS22

dsc07331For industrial applications the reproducibility of spectral measurements and sturdiness and durability of the measuring device is crucial. For reproducibility of spectral measurements between different devices and within the lifetime of the spectrometer, the slit and the position of the grating, the mirror and the detector must be the same. Only this way spectral evaluations, e.g. for the determination of certain analyte concentrations, can be transferred to equivalent spectrometers.

The raspberry spegg fulfills these requirements. Because of the monolithic setup the optical path is grounded over the entire lifetime. Each spectrometer is calibrated after production and does not need to be recalibrated. This manufacturing process ensures that the spectra of all raspberry spegg’s are comparable to each other. For each raspberry spegg, the calibration data is stored in the flash C of the microcontroller.

The electronic components of the raspberry spegg, including the integrated LED, also meet these requirements. E.g. no electrolytic capacitors with liquid electrolytes are used. The light source has its own electronically adjustable voltage source, and the current can be adjusted up to 200mA.

The electronics and the spectrometer of the raspberry spegg are adjusted robustly and space-saving to the raspberry pi.

Only four GPIO’s (pin 4,6,8 and 10) are used for power supply and communication. All other GPIO’s as well as the CSI Camera cable can be used with the raspberry spegg board.

The raspberry spegg has a voltage supply connection with a polyfuse, which can also be used to supply the raspberry pi.

For more pictures click here. For more information here.

spegg-v22-ohne-display

spegg © VIS22

Art.No.: LIVIS2201

EAN: 791732154634

specifications:

Spectral Range

350 – 850 nm (330 – 1050 nm accessible)

Fiber

Ø 300 / 330 µm; NA= 0,22

Fiber length from spegg board to top

35cm +- 1cm

Fiber connector

male F-sma 905

Blazed Grating

g = 4,7 µm / d = 0,2 µm

Stray light Attenuation

> 17dB with GG495 at 470 nm (30 dB using SC30)

Spectral Resolution

< 10 nm FWHM / 3,5 nmPixel

Spectral Accuracy

2 nm (typ. 1 nm)

Inter Instrument Agr.

typ. 0,5 nm

Wavelength Stability

0,05 nm/K

Wavelength Repeatability (RT)

0,05 nm

Sensitivity (e.g. 500nm)

> 15000 cs*m^2 / W*ms

Noise Equivalent Power

3 pW 500nm; Ti 1s

Dynamic Range

> 5.000

Electronics

16 bit; UART;

Wavelength range LED

420 – 750 nm

Light source connector

female F-sma 905

Storage Temperature

– 40°C to +60°C

Operating Temperature

0°C to +40°C

Max. rel. humidity (not condensing)

85 %